Thermocouple – measuring inserts (data sheet 8661)
To order, contact us online, chat, or call (780) 463-9527 Monday-Friday 7.30am-5pm.
Original price
$0.00
-
Original price
$0.00
Original price
$0.00
$0.00
-
$0.00
Current price
$0.00
Thermocouple – measuring inserts (data sheet 8661)
| model | TTeMiAo / TTeMiAoT |
|---|---|
| measuring element | K, N, J and S according to DIN EN 60 584 |
| operating temperature range | type K, NiCr-Ni −40 / +1175 °C (−40 / +2147 °F) type N, NiCrSi-NiSi −40 / +1175 °C (−40 / +2147 °F) type J, Fe-CuNi −40 / +750 °C (−40 / +1382 °F) type S, Pt10Rh-Pt 0 / +1175 °C (+32 / +2147 °F) |
| ambient temperature range | TTeMiAo: −40 / +100 °C (−40 / +212 °F) TTeMiAoT: −40 / +85 °C (−40 / +185 °F) |
| accuracy | class 1 according to DIN EN 60 584 |
| degree of protection | IP 00 measuring inserts are intended for the installation into protective fittings for electrical thermometers, which are equipped with an appropriate degree of protection for safe operation |
| output signal | model TTeMiAo: thermoelectric voltage acc. to DIN EN 60 584 model TTeMiAoT: 4...20 mA, HART® or PROFIBUS® PA / FOUNDATION™ Fieldbus |
| hot junction | hot junction insulated |